Calibration— Type 323
g. ADJUST—C446 (see Fig. 6-17B) for optimum linearity
mid liming between flic sixth arid tenth vertical lines.
h. Set the time-mark generator for 0.5-microsecond mark
ers.
i. Set the TIME/DIV switch to 5
/us.
j. Set the TRIGGER control for a stable display.
k. CHECK— CRT display for optimum linearity and timing
between the second and tenth vertical lines (see Fig. 6-17C).
l
. ADJUST—C454 (see Fig. 6-17B) for optimum linearity
and timing between the second and tenth vertical lines.
m. CHECK— CRT display for optimum linearity and timing
between the left (first) vertical line of the graticule and the
fourth vertical line (see Fig. 6-17C).
n. ADJUST— C432 (see Fig. 6-17B) for optimum linearity
and timing between the left (first) vertical line of the grati
cule and the fourth vertical line.
o. Repeat parts c through n of this step until optimum
timing and linearity is obtained.
37. Check Normal Sweep Timing Accuracy
a. Change the following control settings:
TRIGGER
Stable positive-slope
triggering
X10 HORIZ MAG
Pushed in
TABLE 6-4
Normal Sweep Timing Accuracy
TIME/DIV
switch
setting
Time-mark
generator
output
CRT display
( m arkers/
division)
A llow a ble
error fo r
given
accuracy
5
fxs
5 microsceond
1
0.24 division
1 0 yAS
10 microsecond
1
(within 3%)
20
fis
10 microsecond
2
50
[is
50 microsecond
1
.1 ms
0.1 millisecond
1
.2 ms
0.1 millisecond
2
.5 ms
0.5 millisecond
1
1 ms
1 millisecond
1
2 ms
1 millisecond
2
5 ms
5 millisecond
1
10 ms
10 millisecond
1
20 ms
10 millisecond
2
50 ms
50 millisecond
1
.1 s
0.1 second
1
,2 s
0.1 second
2
.5 s
0.5 second
1
0.32 division
1 s
1 second
1
(within 4%)
b.
CHECK— Using the TIME/DIV switch and time-mark
generator settings given in Table
6-4,
check normal sweep
timing within the given tolerances over the middle eight divi
sions of the display. Set the TRIGGER control as necessary
for a stable display in the variable positive-slope area.
38. Check Magnified Sweep Timing Accuracy
a. Change the following control settings:
TRIGGER
Stable positive-slope
triggering
X10 HORIZ MAG
Pulled out
b. CHECK— Using the TIME/DIV switch and time-mark
generator settings given in Table 6-5, check magnified tim
ing within the given tolerances over the middle eight-division
portion of the total magnified sweep length. Set the TRIG
GER control as necessary for a stable display in the variable
positive-slope area.
c. Disconnect all test equipment and remove the calibra
tion shield.
TABLE 6-5
M agnified Sweep Timing Accuracy
TIME/DIV
switch
setting
Time-mark
generator
output
CRT display
( m arkers/
division)
A llow a ble
error for
given
accuracy
5 /x s
0.5 microsecond
1
0.4 division
10
[is
1 microsecond
1
(within 5%)
20
[is
1 microsecond
2
0.32 division
50
[is
5 microsecond
1
(within 4%)
.1 ms
10 microsecond
1
.2 ms
10 microsecond
2
.5 ms
50 microsecond
1
1 ms
0.1 millisecond
1
2 ms
0.1 millisecond
2
5 ms
0.5 millisecond
1
10 ms
1 millisecond
1
20 ms
1 millisecond
2
50 ms
5 millisecond
1
.1 s
10 millisecond
1
,2 s
10 millisecond
2
.5 s
50 millisecond
1
0.4 division
1 s
0.1 second
1
(within 5%)
6 - 2 2
Summary of Contents for 323
Page 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Page 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Page 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Page 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Page 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Page 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Page 157: ...BL OCK DIAGRAM ...
Page 158: ......
Page 161: ...A TYPE 323 OSCILLOSCOPE ...
Page 162: ...1 TYPE 323 OSCILLOSCOPE ...
Page 163: ...FIG 2 CABINET ...
Page 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...