Appendix A
Table A-1
AUTO-SETUP C ontrol S ettings
FRONT-PANEL CONTROL
AUTO-SETUP ACTION
SCOPE BW
Off.
X I0 MAG
Off.
Horizontal POSITION
Midrange, if not within 1 division of
midrange.
The Horizontal position control does not
affect the position of the trace until moved
through the point corresponding to the
trace position set by AUTO SETUP.
Horizontal MODE
A, if A or X-Y selected; ALT if ALT or
B selected.
TRIGGER
A/B SELECT
A, if A Horizontal MODE selected by
AUTO-SETUP; otherwise B.
A MODE
AUTO LEVEL.
B MODE
RUNS AFTER, if ALT or B selected;
otherwise as selected.
A SOURCE
VERT.
B SOURCE
VERT, if ALT or B selected; otherwise
as selected.
A CPLG
NOISE REJ, if SOURCE Volts/Div is set to
2 mV by AUTO SETUP, otherwise DC.
B CPLG
NOISE REJ, if SOURCE Volts/Div is set to
2 mV by AUTO SETUP, otherwise DC.
A SLOPE
As selected.
B SLOPE
As selected.
A LEVEL
Nominally midway between peaks of A
SOURCE signal. Signal peaks must be
< (10 volts X probe multiplier).
B LEVEL
Nominally midway between peaks of B
SOURCE signal. Signal peaks must be
< (10 volts X probe multiplier).
CH 1 and CH 2 COUPLING
(Channel turned on)
DC, if COUPLING set to GND or DC;
otherwise AC.
2245A Operators
A-1