Performance Check Procedure
l
.
Move sine-wave generator signal to the CH 4 Input.
m. Set:
CH 1, CH 2, and CH 3
VERTICAL MODE
On (CH
4 off)
TRIGGER SOURCE
CH 4
n.
CHECK—display amplitude Is 0.1 division or less, excluding trace
width, on the CH 1, CH 2, and CH 3 traces.
o.
Disconnect the test setup.
21. AC-Coupled Lower -3 dB Point
a.
Set:
A SEC/DIV
VERTICAL MODE
TRIGGER SOURCE
TRIGGER MODE
10 ms
CH 1 (all others off)
VERT
NORM
b.
Connect function generator (FG 502) output to the CH 1 input via a
50 f l BNC coaxial cable and a 50 f l BNC termination.
c.
Set the function generator output controls to produce a six-division
sine-wave display at 10 Hz (with no dc offset).
d.
Set CH 1 Input COUPLING to AC.
e.
CHECK—display amplitude is 4.2 divisions or more.
f.
Set VERTICAL MODE to CH 2 (CH 1 off).
g.
Repeat the procedure for CH 2.
h.
Disconnect the test equipment from the 2245A.
2245A Operators
7-25