![Tektronix 2245A Operator'S Manual Download Page 133](http://html.mh-extra.com/html/tektronix/2245a/2245a_operators-manual_1078849133.webp)
Performance Check Procedure
e.
CHECK—that the sine-wave generator output frequency is between 17
MHz and 23 MHz.
f.
Disconnect the test setup.
19. Common-Mode Rejection Ratio
a.
Connect leveled sine-wave generator (SG 503) output to the CH 1 and
CH 2 input connectors via a 50
Cl
precision BNC coaxial cable, a 50 f 1
BNC termination, and a BNC dual-input coupler.
b.
Set the leveled sine-wave generator output for an eight-division signal-
display amplitude at 50 kHz.
c.
Set:
VERTICAL MODE
CH 2 VOLTS/DIV
CH 2 INVERT
SCOPE BW
ADD (CH 1 off)
10 mV
On
Off
d.
Adjust CH 1 or CH 2 VAR VOLTS/DIV (as needed) for smallest signal
amplitude.
e.
Set the leveled sine-wave output frequency to 50 MHz.
f.
Set the VERTICAL MODE to CH 1 (ADD off).
g.
Set the leveled sine-wave output amplitude for an eight-division dis
play.
h.
Set the VERTICAL MODE to ADD (CH 1 off).
i.
CHECK—the signal is less than 0.8 division in amplitude.
j.
Disconnect the test setup.
2245A Operators