3.3.2
Transmitter Mounting
The Tek-Cor 1100A can be mounted directly or remotely with the help of a nine-core
cable.
Install the sensor and the transmitter in the pipeline. The transmitter can be mounted
with 90° rotation according to the requirement of sensor installation as shown in the
picture above.
3.4
Wiring
3.4.1
Basic Requirements
If the sensor of the Tek-Cor 1100A Coriolis Mass Flow meter is installed integrally with
the transmitter, it is easy to power up the transmitter. If the sensor of Tek-Cor 1100A
is installed separately from the transmitter then the transmitter is connected with the
sensor through special nine-core cable.
3.4.2
Junction Box
If the sensor and the transmitter are installed separately, ensure that the sensor and
transmitter have been respectively matched with junction box for connecting the
special nine-core cable.
CAUTION
Cut off power before connecting the cables. The power input voltage must match the
voltage specified in the junction box of the transmitter. The earth connector must be well
connected with earth wire to ensure its intrinsic safety performance.
3.4.3
Cable Connection
A nine-core cable is used when the sensor and the transmitter are connected
separately. The following picture shows a nine-core cable.