12
2.4.3
Flow Range for Gas
Size (Inch)
Measurable Flow Range
(lb/min)
Flow Range with 0.5%
(lb/min)
1/2”
1 - 110
3 - 110
1”
2 - 293
8 - 293
1 ½”
12 - 1176
30 - 1176
2”
19 - 1837
46 - 1837
3”
26 - 5144
129 - 5144
4”
37- 7348
184 - 7348
6”
92 - 18371
460 - 18371
2.4.4
Flow Range of Volume of Air under STP Conditions
Size (Inch)
Initial flow
(GPM)
Initial flow
(Scfm)
Flow range with
accuracy 0.5%
(GPM)
Flow range with
accuracy 0.5%
(Scfm)
1/2”
55
8
275~11007.5
40 – 1585
1”
146.5
21
734~29353
106 – 4226
1 ½”
587
84
2935.5~117411
423 – 16906
2”
917.5
132
4586.5~183454.5
661 – 26416
3”
2568.5
370
12842~513672
1849 - 73966
4”
3669
528
18345.5~733817
2642 - 105665
6”
9172.5
1321
45863.5~1834542
6604 – 264164
2.4.5
Flow Rate Factor
The flow rate of the Tek-Cor 1100A must be checked according to the formula below:
Medium flow rate =
𝑽𝑽𝑽𝑽𝑽𝑽𝑽𝑽𝑽𝑽𝑽𝑽
𝒇𝒇𝑽𝑽𝑽𝑽𝒇𝒇
𝒓𝒓𝒓𝒓𝒓𝒓𝑽𝑽
𝑽𝑽𝒖𝒖𝒖𝒖𝑽𝑽𝒓𝒓
𝒇𝒇𝑽𝑽𝒓𝒓𝒘𝒘𝒘𝒘𝒖𝒖𝒘𝒘
𝒄𝒄𝑽𝑽𝒖𝒖𝒖𝒖𝒘𝒘𝒓𝒓𝒘𝒘𝑽𝑽𝒖𝒖𝒄𝒄
𝑭𝑭𝑽𝑽𝑽𝑽𝒇𝒇
𝒓𝒓𝒓𝒓𝒓𝒓𝑽𝑽
𝒇𝒇𝒓𝒓𝒄𝒄𝒓𝒓𝑽𝑽𝒓𝒓
Size (inch)
½”
1”
1 ½”
2”
3”
4”
6”
8”
Flow rate
factor
0.362
1.558
1.634
3.535
8.338
15.89
26.15
58.84
NOTE
•
The gas flow velocity is usually much higher than liquid when measured by a flow
meter, so there is noise caused by gas medium and tube wall of flow meter
under a high-speed gas flow and if the noise becomes loud enough, the signal of
flow meter will be influenced, so use the Tek-Cor 1100A Coriolis flow meter for
gaseous medium at velocity less than 1/3
rd
of the sound velocity (360 fps).
•
Use Tek-Cor 1100A for gas with a pressure less than 29 PSI (0.2 MPa).