5.3.2 SEDATION MENU
Screen description:
1 - PATIENT
2 - TOTAL FLOW (by pressing on the total flow it is possible to display the individual flows of O
2
and N
2
O
in position 4)
3 - N
2
O / O
2
PERCENTAGE (pressing it displays the O
2
and N
2
O % alternately)
4 - O
2
/ N
2
O FLOW
5 - AUTONOMY, the available volume in NORMAL LITRES is displayed. If the device is delivering, the
remaining time in MINUTES is displayed. It is possible to choose whether to display the autonomy, using
the appropriate command (press menu
view and select the appropriate item).
6 - VACUUM, the SUCTION FLOW is displayed. It is possible to choose whether to display the suction
flow by using the appropriate command (press menu
display and select the appropriate item). The
flow is only displayed when using the active suction circuit.
7 - BASELINE the baseline values (total flow and N
2
O %) are displayed, if previously set. To set them
press the Save Baseline icon during sedation, this saves the flow of O
2
and N
2
O that the device is
delivering. It is also possible to set/change it directly on the patient form. It is not possible to set O
2
values lower than 30%.
8 - % N
2
O / % O
2
BAR is a direct display of the percentage of gas delivered.
9 - DISPLAY / RECORDING (red dot when recording).
Recording starts when more than 0.5 nl/min starts to be delivered and ends when gas
delivery drops below 0.5 nl/min.