15. Connect the device pin to a socket with a rated voltage of 230V
±
10% MON EARTH,
frequency 50/60Hz, current 12A, as shown on the identification plate place in the rear side of
the device.
16. Never connect the device pin to reductions of any type.
17. Load the filling tank as explained on PAR. 8 of this manual.
18. After the installation, carry out a LOADED sterilisation cycle as explained on PAR. 10 of this
manual.
Photo A
6. INTRODUCTION FOR STERILISATION
The instruments to be sterilised must be previously cleaned, free from residual materials, blood,
deposits, etc. The above mentioned rests can damage the materials to be sterilised or the
autoclave components; moreover, they do not ensure a perfect sterilisation.
1. Wash the instruments after using in cold water to remove any rest.
2. Insert the instruments in the ultrasound machine.
3. Rinse the instruments with clean water with care.
4. Dry the instruments to avoid that not demineralized water remains on them, this causing
oxidation or instruments spots due to the calcareous rests.
5. Wash, rinse and dry also the supplied trays.
6. On the same tray only instruments of the same material must be placed.
7. The small mirrors must be positioned with glass downwards.
8. For a better sterilisation open the instruments such as forceps, shears or other composed
instruments.
9. Do not stack the instruments on the trays. The instruments must be sterilised separately. An
overload could compromise the sterilisation.
10. If empty containers are sterilised to avoid a water accumulation, it is necessary to place them
upside down.
11. Sterilise the instruments only on the trays inserted in the supplied tray - holder; there must be a
certain distance between the trays so that vapour circulates during the sterilisation phase, this
easing drying.
Summary of Contents for Hydra Evo
Page 5: ......
Page 6: ......
Page 24: ...Foto F Agire in senso orario per ottenere l effetto opposto...
Page 26: ......
Page 38: ...Photo F Turn it clockwise to decrease the pressure...
Page 40: ......
Page 52: ...Foto F Tourner droite pour avoir l effet contraire...
Page 67: ......
Page 79: ...Foto F Durch Rechtrumdrehen wird der gegens tzliche Effekt erzielt...
Page 92: ...P Figura F Rode no sentido dos ponteiros do rel gio para diminuir a press o...
Page 94: ...P...
Page 101: ......