ENGLISH
18
05
PRINCIPLE OF OPERATION
The following describe an
OPERATING CYCLE
and a
VACUUM TEST CYCLE
(refer to
the user manual for the timing of each cycle that the autoclave can make).
PREHEATING
The autoclave is equipped with a fully automatic preheating function that speeds up the
operating cycles of the autoclave.
After having installed the machine and removed the accessory kit form the chamber, turn
on the device moving the button on the side of the control panel
(X)
in position – I – and
open the door.
When the autoclave turns on the display shows the name of the machine and then a check
of the devices status with the firmware release.
After a few seconds the displays show the main screen of the machine followed by a beep,
indicating the effective barometric alignment.
The preheating is triggered in all operating and test cycles
with the exception of the
VACUUM TEST
where it will remain switched off. In this phase the resistance brings the
temperature of the chamber surface at a temperature of 80°.
Prepare the instruments to be sterilized, place them on the appropriate tray and on the tray
holder seats and insert everything inside the chamber and close the door manually by
turning the close lever.
CAUTION BURN HAZARD
Autoclave in pre-heating
Select the desired cycle by pressing
SELECT
to scroll the cycles.
When the desired cycle is highlighted, press the
Start
key in correspondence of the
multifunction keys to start the cycle.
All the sterilisation cycles involve 3 vacuum stages except for the Safety Cycle that
only features 2 vacuum stages, switching from vacuum 1 to vacuum 3.
Summary of Contents for ONYX B 5.0
Page 2: ......
Page 11: ...ITALIANO 11 ONYX 8 0 EV5 EV3 EV7 EV4 EV2 EV6 FLOWMETER PRESSURE TRANSDUCER...
Page 16: ...ITALIANO 16 COLLEGAMENTO BASE...
Page 17: ...ITALIANO 17 COLLEGAMENTO CON OSMOSI COLLEGAMENTO CON SISTEMA OSMOSI CENTRALIZZATO...
Page 71: ...ITALIANO 71 MAIN BOARD...
Page 72: ...ITALIANO 72 REFERENCE TYPE CONNECTOR PIN NUMBERS N SIGNAL NAME...
Page 76: ...ITALIANO 76 USER BOARD...
Page 88: ...ENGLISH 11 ONYX 8 0 EV5 EV3 EV7 EV4 EV2 EV6 FLOWMETER PRESSURE TRANSDUCER...
Page 93: ...ENGLISH 16 BASIC CONNECTION...
Page 94: ...ENGLISH 17 PURA PLUS OSMOSIS SYSTEM INTALLATION CENTRALIZED OSMOSIS SYSTEM INTALLATION...
Page 148: ...ENGLISH 71 B ELECTRIC AND CONNECTION DIAGRAM...
Page 149: ...ENGLISH 72 MAIN BOARD...
Page 153: ...ENGLISH 76 USER BOARD...
Page 165: ...ESPA OL 11 ONYX 8 0 EV5 EV3 EV7 EV4 EV2 EV6 FLOWMETER PRESSURE TRANSDUCER...
Page 170: ...ESPA OL 16 INSTALACI N B SICA...
Page 171: ...ESPA OL 17 INSTALACI N CON EL SISTEMA DE SMOSIS INSTALACI N CON SISTEMA DE SMOSIS CENTRALIZADO...
Page 225: ...ESPA OL 71 B ESQUEMA EL CTRICO Y CONEXIONES...
Page 226: ...ESPA OL 72 MAIN BOARD...
Page 230: ...ESPA OL 76 USER BOARD...
Page 234: ......