8. Maintenance and Cleaning
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
69
8.6.3
Installing the Manifold
WARNING
ALL PARTS OF THE INSTRUMENT THAT COME INTO CONTACT
WITH POTENTIALLY INFECTIOUS MATERIAL MUST BE TREATED AS
POTENTIALLY INFECTIOUS AREAS.
IT IS ADVISABLE TO ADHERE TO APPLICABLE SAFETY
PRECAUTIONS, (INCLUDING THE WEARING OF POWDER-FREE
GLOVES, SAFETY GLASSES AND PROTECTIVE CLOTHING) TO
AVOID POTENTIAL INFECTIOUS DISEASE CONTAMINATION WHEN
PERFORMING CLEANING PROCEDURES AND ALSO WHEN MAKING
ADJUSTMENTS TO THE INSTRUMENT.
The standard manifolds are installed using the following procedure:
1. Lift the manifold arm.
2. Carefully fit the manifold onto the manifold arm and ensure that the locating
pins are correctly inserted through the holes in the bracket.
3. Tighten the manifold into place using the two screws using the Allen key
provided.
4. Fit the dispensing tubes (marked blue) onto the top connector on the rear of
the manifold.
5. Fit the aspirating tubes (marked red) onto the two bottom connectors on the
rear of the manifold (marked with red labels).
Locating Pin
Locating Pin
2 screws
3 screws