9. Trouble Shooting and Error Messages
2022-01
HYDROFLEX PLUS No. 30199906 Rev. No. 1.1
77
9. Trouble Shooting and
Error Messages
9.1
Errors
The following errors can occur, which will not yield an error message from the
standard
instrument’s firmware:
Error Description
Possible Causes
See Chapter
Display and LED dark
Fuse is defective.
Contact your local Service
Representative
No or incorrect dispensing Dispensing needles
blocked.
8.3.2 Cleaning the Liquid
System
No or incorrect aspirating Aspirating needles
blocked.
8.3.2 Cleaning the Liquid
System
9.2
Error Messages
9.2.1
Standard Instrument: Error Messages
Plate Error
If the plate sensor does not recognize the plate or no plate is inserted on the plate
carrier, the following message is displayed:
NO PLATE
OK
Press
OK
and insert the plate correctly on the plate carrier
Program Too Big Error
If the program has been defined with more than 60 process steps (each cycle is
worth 2 steps), the following message is displayed:
PROGR: TOO BIG
OK
Press
OK
and define a program with less than 60 process steps.
Transport Error
If the instrument cannot move the plate carrier, the following message is
displayed:
INIT ERROR
OK
Press
OK
to remove the error message and the instrument returns to the standby
mode.
Check that the plate carrier is correctly inserted and that the plate carrier system
is clear.