28
Group
Setting name
Command
No.
R/W
Default value
(actual unit)
Command data (setting
value)
Operation
explanation
D/A Converter*
D/A Output Mode
1301
R/W
0: Voltage
0: Voltage
1: Current
D/A Maximum Voltage
1302
R/W
10 V
1 V − 10 V
D/A Zero
1303
R/W
00000
(000.00)
−99999 to 99999
D/A Full Scale
1304
R/W
10000
(100.00)
−99999 to 99999
Display
Minimum Scale Selection
1402
R/W
0: 1
0: 1
1: 2
2: 5
3: 10
Display Frequency Selection
1403
R/W
4 times
0: 4 times
1: 6 times
2: 10 times
3: 20 times
Maximum Display Value
1404
R/W
11000
(110.00)
00000 to 99999
Sensor Input Logic
1405
R/W
0: Standard
0: Standard
1: Reversed
*The D/A converter setting only affects D/A models.
6-6-4. Condition settings
Group
Setting name
Command
No.
R/W
Default value
(actual unit)
Command data (setting
value)
Operation
explanation
Filter
Low-Pass Filter Selection
2001
R/W
4: 100 Hz
0: OFF
1: 3 Hz
2: 10 Hz
3: 30 Hz
4: 100 Hz
5: 300 Hz
6: 1000 Hz
Moving Average Data Set
Selection*
2002
R/W
0: OFF
0: OFF
1: 16 times
2: 32 times
3: 64 times
4: 128 times
5: 256 times
6: 512 times
7: 1024 times
8: 2048 times
Automatic Digital Filter
2003
R/W
1: ON
0: OFF
1: ON
Moving Average Data Set Size*
2004
R/W
0: OFF
0: OFF
1: Disabled
2–2048: Moving average data
set size
Motion Detect
Time
2101
R/W
15
(1.5 second)
0 to 99
(changed to 0.0 to 9.9 seconds
internally)
Width
2102
R/W
00005
(000.05)
00000 to 00999
Zero Tracking
Time
2201
R/W
0
(0.0)
0 to 99 (changed to 0.0 to
9.9 seconds internally)
Width
2202
R/W
00000
(000.00)
00000 to 00999
6. RS-485