43
9-2. Motion Detect
Set the parameters used to detect stability.
The indicator value is judged to be stable if a difference less than the set width occurs between the current indicator value and the
indicator value 100 mSec before, and this continues for the set time.
The motion detect function is closely related to the comparison mode. See “10-3. Comparison Mode Setting” for details.
Indicator value
Motion detect
width
Motion detect time
Motion detect time
100 msec
100 msec
100 msec
Time
Stable
Stable
Stable
judgment
Stable
judgment
Not stable
judgment
Not stable
9-3. Zero tracking
This function automatically tracks drift and other gradual changes to the zero point.
With Zero tracking, when the zero point moves no more than the set width, the indicator value is automatically adjusted to zero each
set time interval.
This only functions when Calibration Value Lock is ON and Enable Digital Zero is ON.
Zero tracking will not function if its time is set to 0.0 seconds and its width is set to 0.0.
0
Zero tracking width
Zero tracking time
Zero tracking active
Zero Tracking active
Indicator value
Time
9. Condition settings