15
3-3-3. Comparison output terminals
Terminal
number
Signal
Explanation
14
HI
High limit judgment output
15
LO
Low limit judgment output
16
OK
OK judgment output
18
−(0 V)
Comparison output common
terminal
o
The comparison output is isolated from the unit’s circuits by
a photocoupler.
Open collector output (NPN, current sync)
20mA/30V maximum collector current
+5V
18 −(0V)
14–16 judgment
output
20 mA/30V
maximum
o
For the operation of each signal, see “10. Comparison set-
tings” on page 45.
3-4. Connecting the DC power
supply input terminals
The DC power supply input voltage should be 24 V ±10%.
Use a power supply cord that is 0.517 to 2.081 mm
2
(20 to
14 AWG).
When connecting to the terminal bank, use a solderless termi-
nal (M3, width of 6 mm or less).
+
–
FG
ATTENTION
o
If using the unit as a CE compliant product, the power cord
must be no more than 3m long.
o
Allow the unit to warm up for at least five minutes after
supplying power to it.
o
The recommended tightening torque for terminals is 0.5 N·m
(5.1 kgf-cm).
V
Always attach the cover to the power supply input
terminal bank. Failure to do so could result in fire,
electric shock or malfunction.
V
Use a power supply that conforms to a limited energy
circuit as specified in IEC61010-1. Failure to do so
could cause fire or other hazards.
3. Making connections