27
6-6. Commands
6-6-1. Execution
Setting name
Command
No.
R/W Default value
Command data (setting
value)
Operation
explanation
Digital Zero
0000
W
10
Clear Digital Zero
0000
W
11
Hold On
0000
W
12
Hold Off
0000
W
13
Hold Lock
0000
W
14
Hold Clear
0000
W
15
Switch to Home Screen
0000
W
17
Static Strain Display Switching
(static strain display mode)
0000
W
20
6-6-2. Polling
Setting name
Command
No.
R/W Default value
Command data (setting
value)
Operation
explanation
Polling
0001
R
Status Polling
0002
R
Indicator Value Polling
0003
R
6-6-3. Calibration
Group
Setting name
Command
No.
R/W
Default value
(actual unit)
Command data (setting
value)
Operation
explanation
Bridge Voltage
1001
R/W
2 (5 V)
1: 10 V
2: 5 V
Decimal Point Position
1002
R/W
2 (0.00)
0: None/1:0.0/2: 0.00/3:
0.000/4: 0.0000
Zero Point Input Calibration
1003
R/W
0 (0 mV/V)
−5000 to 5000
(changed to −5.000 to
5.000 mV/V internally)
Zero Balancing
1004
R/W
0
When W, command data will
be ignored and zero balanc-
ing will be conducted.
When R, the current setting
value will be returned as μV/V.
Remote Sense
1005
R/W
0
0: Unused (4-wire format)
1: Used (6-wire format)
Equivalent Input
Calibration
Rated Output Value
1101
R/W
3000
(3.000 mV/V)
50 to 5000
(changed to 0.050 to
5.000 mV/V internally)
Rated Capacity Value
1102
R/W
10000
(100.00)
00001 to 99999
Actual Load Calibration Rated Capacity Value
1103
R/W
10000
(100.00)
00001 to 99999
TEDS Calibration
TEDS Calibration
1104
W
6. RS-485