In this chapter we explain the D/A converter functions of the
TD-SC1 (D/A).
The D/A converter settings only affect TD-SC1 (D/A) models.
The D/A converter allows for analog output that corresponds
to the unit indicator value.
The analog output range is either 0–±10V voltage output or
4–20mA current output. Set the maximum voltage output
between ±1 V and ±10 V in 1V steps using the D/A Maximum
Voltage setting.
Analog output from zero (0 V, 4 mA) to full scale (±10 V, 20 mA)
can be achieved according to the digital values set using D/A
Zero and D/A Full Scale settings.
The zero point and full scale cannot be changed separately for
current output and voltage output.
7-1. D/A output terminals
These can output either voltage or current.
They cannot output both at the same time. Use output mode
commands to set the voltage and current (page 34).
Signal name
Explanation
SLD
GND with external device
V-OUT
D/A voltage output
I-OUT
D/A current output
COM
D/A output common terminal
o
The D/A output is isolated from this unit’s circuits.
For voltage output, connect an external device with a load
resistance of 2 kΩ or more to V-OUT and COM.
For current output, connect an external device with a load
resistance, including cable wiring resistance, of 350 Ω or less to
I-OUT and COM.
NOTE
With static strain display, analog output will be zero.
External device
Load resistance
2 kΩ or more
Shield
Shield
External device
Load resistance
350 Ω or less
See “7-1. D/A output terminals” on page 33 for terminal
numbers.
7-2. D/A Zero
Set the indicator value to output for D/A Zero (0V voltage or
4mA current).
7. D/A converter