7-3. D/A Full Scale
With the “7-2. D/A Zero” value as the reference, set the span for
the indicator values output to D/A.
When the “7-2. D/A Zero” and “7-3. D/A Full Scale” setting values
are added, the “7-5. D/A Maximum Voltage” setting value volt-
age (20 mA when in electrical current mode) is output.
The following are examples of when “7-5. D/A Maximum
Voltage” is set to 10 V.
Setting example 1
D/A Zero
000.00
D/A Full Scale
100.00
100.00
10 V (20 mA)
0.00
0 V (4 mA)
−100.00
−10 V(-- mA)
Indicator value
D/A output
Setting example 2
D/A Zero
020.00
D/A Full Scale
100.00
120.00
10 V (20 mA)
20.00
0 V (4 mA)
−80.00
−10 V (--mA)
Indicator value
D/A output
Setting example 3
D/A Zero
020.00
D/A Full Scale
−100.00
120.00
−10 V (-- mA)
20.00
0 V (4 mA)
−80.00
10 V (20 mA)
Indicator value
D/A output
Setting example 4
D/A Zero
−010.00
D/A Full Scale
020.00
10.00
10 V (20 mA)
−10.00
0 V (4 mA)
−30.00
−10 V (--mA)
Indicator value
D/A output
ATTENTION
If calibration is conducted, when the rated capacity is con-
firmed, it will be set as the D/A full scale value.
7-4. D/A Output Mode
Select voltage or current.
7-5. D/A Maximum Voltage
Voltage setting (limiter)
o
The output has an extended range of about 10% more than
the set value. The output voltage range is the same in the
negative direction. For example, when set to 5 V, the D/A
output voltage range will be about −5.5 V to +5.5 V, and
+5 V will be output when the “7-2. D/A Zero” and “7-3. D/A
Full Scale” setting values are added.
7. D/A converter