GLOBAL COMMANDS
The following global commands affect all power supplies on a multi-drop chain. See
section 9.4.2 for requirements. There is no query version of these commands
9.4.4. Global Set the Voltage Limit
This is the global version of the
:VOLT
command in section 7.2.1
Syntax:
GLOBal:VOLTage[:LEVel][:IMMediate][:AMPLitude] <nn.nn>
Parameter:
<nn.nn>
is volts, the setting applied to all power supplies
Example:
GLOB:VOLT 9.45
9.4.5. Global Set the Current Limit
This is the global version of the
:CURR
command in section 7.2.2
Syntax:
GLOBal:CURRent[:LEVel][:IMMediate][:AMPLitude] <nn.nn>
Parameter:
<nn.nn>
is amps, the setting applied to all power supplies
Example:
GLOB:CURR 350
9.4.6. Global Set the Output ON or OFF
This is the global version of the
OUTP:STAT
command in section 7.2.3
Syntax:
GLOBal:OUTPut:STATe <0|1|OFF|ON>
Example:
GLOB:OUTP:STAT 1
(or
GLOB:OUTP:STAT ON
)
9.4.7. Global Reset Power Supplies
This is the global version of the
*RST
command in section 7.7.2
Syntax:
GLOBal:*RST
Example:
GLOB:*RST
9.4.8. Global Save All Settings
This is the global version of the
*SAV 0
command in section 7.7.3
Syntax:
GLOBal:*SAV 0
Example:
GLOB:*SAV 0
9.4.9. Global Recall All Settings
This is the global version of the
*RCL 0
command in section 7.7.4
Syntax:
GLOBal:*RCL 0
Example:
GLOB:*RCL 0
9.4.10. Error and Status Registers
The error and status registers described in sections 7.8.5 through 7.8.17 operate somewhat
differently in a multi-drop system.