*TST? Alternate form of IST? query.
AV1
Amps/Volt Pulsing Base Level
Description
This command is used to set the base level when pulse loading in
the Amps/Volt mode.
NOTE:
AVL or AVH must have been previously set.
Command Syntax
AV1 <NR2>
Example
AV1 25.0
This will set the base Amps/Volt level to 25 amps/volt
Query Syntax
AV1?
Returned Parameters
TEXT ON: <NR2> amps/v
TEXT OFF: <NR2>
AV2 Amps/Volt Pulsing Peak Level
Description
This command is used to set the peak level relative to the base level
when pulse loading in the Amps/Volt mode.
NOTE:
AVL or AVH must have been previously set.
Command Syntax
AV2 <NR2>
Example
AV2 50.0
This will set the peak level to 50 Amps per Volt above the base
level. Therefore if the base level setting is 25A, the peaks will be 75
Amps per Volt.
Query Syntax
AV2?
Returned Parameters
TEXT ON: <NR2> amps/v
TEXT OFF: <NR2>
Page 52 Of 162