SHORT
Short Circuit
Description
This command is used to apply or remove a simulated short circuit
from the device being tested while the load connect relay is closed.
CAUTION:
This function is intended for use only when testing current limited power
sources. Use of the Short Circuit feature with power sources which do not limit
their current to values within the load’s ratings could result in damage to the
load or device under test,
FIRE or EXPLOSION.
Command Syntax
SHORT ON
SHORT OFF
Example
SHORT ON
This will apply a simulated short circuit at the load’s input
terminals.
SHORT OFF
This will remove the simulated short circuit from the load’s input
terminals.
Query Syntax
SHORT?
Returned Parameters
Text
On:
Text
Off:
SHORT ON
SHORT OFF
0 (meaning Off)
1 (meaning On)