Place the blade guard over the blade when sectioning.
Make sure that liquids do not enter the interior of the instrument.
1.2.3 Warnings-Cleaning and Maintenance
Only authorized and qualified service personnel may access the internal
components of the instrument for service and repair.
Before each cleaning, switch the unit off, disconnect the power plug, and
remove the blade holder completely and clean it separately.
Lock the hand wheel before each cleaning.
Do not use any solvents containing acetone or xylene for cleaning.
Make sure that liquids do not enter the interior of the instrument when
cleaning.
Let the unit dry completely before powering up again.
Turn the unit off and disconnect the power plug before replacing fuses. Only
use fuses of the same specification and replace them as described in the
manual.
Safety Devices
Handwheel Locking Mechanism
There are two handwheel locking mechanisms
.
•
Locking mechanism
indicated by (1) on the
diagram can be engaged by flipping the
corresponding switch and rotating the hand
wheel to the upper most position where it will
lock into position.
•
Locking mechanism
indicated by (2) on the
diagram can be engaged in any position of the
handwheel by rotating the handle 180°
clockwise. To unlock, rotate 180° counter
clockwise.
Do not lock the handwheel (1) when the handwheel is rotating as it will
damage the unit.
Whenever cleaning the unit, changing the specimen, or changing the blade,
make sure the handwheel is locked with either mechanism (1) or (2).