SECT Button
•
SECT button activates the sectioning mode. The first line on the
display will show [SECT=]. Used to adjust section thickness
TRIM Button
•
TRIM button activates the trimming mode. The first line on the
display will show TRIM=. Used for setting the TRIM thickness
•
When the unit is powered on, it will display the last set value for
the section or trimming mode.
•
Pressing the CLEAR button will clear the section / trimming counter
and the thickness sum (total advance).When the instrument is
energized, the value of the section counter and section thickness sum
will automatically go back to zero.|
•
Press the + (plus) button or – (minus) button to set the section or trim
value.
Adjusting range:
0.5 to 600µm
0.5 to 2µm
,
0.5µm increments
2 to 10µm
,
1µm increment
10 to 20µm
,
2µm increments
20 to 100µm
,
5µm increments
100 to 600µm
,
50µm increments
•
The arrow buttons control the forward and backward direction of the
arm that holds the specimen. The advance and retract speed is
750µm/s.
•
The max forward and backward distance is 20mm. If the maximum
distance is exceeded, the specimen holder will stop moving and an
alarm will sound.