Table 8. Electromagnetic Emissions
The oCare
TM
Pro100 is intended for use in the electromagnetic
environment specified below. The customer or user of the
monitor should assure that it is used in such an environment.
Emission Test
Compliance
Electromagnetic Environment -
Guidance
RF Emissions
CISPR 11
Group 2
This oCare
TM
Pro 100 must
emit electromagnetic energy
in order to perform its
intended function. Nearby
electronic equipment may
be affected.
RF Emissions
CISPR 11
Class B
The oCare
TM
Pro 100 is
suitable for use in
establishments, including
diagnostic establishments
and those directly
connected to the public
low-voltage power supply
network that supplies
buildings used for domestic
purposes.
Harmonic
Emissions.
IEC 61000-3-2
N/A
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
N/A