overwrite low priority alarms.
2
For multiple alarms, lower priority alarms would also be
displayed.
3
Alarm system still operates during adjustment of alarm
limit.
WARNING:
1.
To ensure the alarm limits are appropriate to the monitored
patient, check the alarm setting each time before use.
2.
Do not change the alarm limit before consulting the
medical advice.
3.
Alarm condition (physiological and technical), alarm
limits, date/time of occurrence are saved in non-volatile
memory. The alarm data is kept intact during powered
down.
4.
When the alarm data is full, the device overwrites the
oldest existing alarm data with the new alarm data.
5.
When the battery is in very low state (<25%), the alarm
system is not expected to repeat alarm signals indefinitely.
6.
If a low battery condition occurs, immediately discontinue
patient monitoring and charge the device.