Taiwan Biophotonic Corporation
59
SpO
2
Low
Limit
85~95%, with 1% step size
85%
PR High Limit 75~250bpm, with 5bpm step size
170bpm
PR Low Limit 40~110bpm, with 5bpm step size
40bpm
Note:
1.
The low limit must be set below the high limit setting.
While the high limit is set equal to or below the low
limit, the low limit will automatically be adjusted to the
next setting below the newly entered high limit setting.
2.
When default alarm setting is changed, a decimal point
(.) is displayed after the reading as a reminder.
3.
When PR/SpO
2
high (or low) limit is exceeded,
Medium priority alarm is activated.
The PR/SpO
2
number on the display turns
YELLOW.
Table 5. Technical Alarm – Low Priority
Parameter
Condition
Low Battery
Remaining battery is less than 25%
Poor Signal Condition Poor signal, determined by the
algorithm, for more than 25 seconds
Malfunction
Malfunction detected
NOTE:
1
Alarms are prioritized so that high priority alarms