1.5 Use and keeping of the manual
The present manual is an integral part of the product and must be kept near the device to be easily and speedily con-
sulted. The purpose of the present manual is to give instructions for:
1.
the correct installation of the device;
2.
the safe and efficient operation of the device;
3.
the continued and compliant maintenance of the device.
The device must be used in accordance with the procedures contained in this manual and never for any purpose other
than provided herein. It is further supposed that the directives with regard to the safety of personnel in force in the des-
tination country of the appliance are known and applied at the place of use of the device.
This manual must be kept in a safe place, easily accessible for the personnel; it must also be handled with care.
1.6 The manufacturer’s responsibilities
The device is constructed in accordance with European regulations.
It should be noted that the use of the appliance is limited to the functions for which it was designed and made.
If the autoclave is improperly used, any resulting damage will be the responsibility of the user.
The product is guaranteed for 12 months from the date of purchase.
1.7 Secondary risks
During the normal operating cycle of the appliance, the operator is exposed to certain risks that cannot be totally elimi-
nated, given the very nature of the machine.
1.
Danger of burns (scalds).
When the autoclave has completed its sterilisation cycle and the operator opens the door to remove the sterilised instru-
ments, their temperature will be between 70° C and 85° C:
beware of the risk of burns, allow to cool for at least two
minutes (door open) before removing from the sterilisation chamber.
Again, at the end of the sterilisation cycle, we indicate the other temperatures in order prevent any burns to the opera-
tor:
Inner part of chamber and door = temperature between 70 °C and 85 °C (Fig. 1)
Outer part of casing = 40 °C
To remove the sterilised instruments from the chamber, use the specific handle supplied (App. 9, page 37). When open-
ing the door, do not stand over or in front of it, there is a risk of scalding by escaping steam (F.1).
2.
Danger of accident to the hands.
• When in operation, the cooling fan of the appliance runs continuously. Do not remove the protective outer casing
without switching off the supply. In the event of anomalies, always contact technical assistance.
Fig. 1: Note the arrow s, they indicate
the hot parts of the machine