TAU STERIL
N° 11 - Rev. 3 03/02/20
User Manual
Page 28
FAST B
8
Hydraulic diagram
S
Heat echanger
(vacuum pump cooling)
S.C.
Clean water tank (filling)
V
Heat echanger fan
(exchanger cooling)
S.SC.
Used water tank (drainage)
PH202
Water pump (2) tank filling
(to fill main water tank)
VS
Safety valve
PH201
Water pump (1) fills steam genera-
tor filling
(to fill steam generator, if required)
P
Pressure transducer
PV
Vacuum pump
(to create vacuum during the cycle)
C
Sterilisation chamber
EV1
2-way solenoid valve N.C. 24V d.c.
(protects water pumps 1)
FB
Bacteriological filter
EV4
2-way solenoid valve N.O. 24V d.c.
(water drainage)
VA
Water collection valve
EV5
2-way solenoid valve N.C. 24V d.c.
(protects vacuum pump)
GV
Steam generator
EV2
2-way solenoid valve N.C. 24V d.c.
(steam outlet)
RF
Water filtering connection
EV6
2-way solenoid valve N.C. 24V d.c.
(bacteriological filt intake)
RS1
Connection for draining clean water
EV7
2-way solenoid valve N.O. 24V d.c.
(steam generator outlet)
RS2
Connection for draining used water
S. VAP Steam vent
LEGEND