![TAU STERIL FAST B User Manual Download Page 18](http://html1.mh-extra.com/html/tau-steril/fast-b/fast-b_user-manual_821621018.webp)
TAU STERIL
N° 11 - Rev. 3 03/02/20
User Manual
Page 18
FAST B
To open the
door
Clearly, the equipment is not sterile as it has not been through the PR Sterilisation phase: the load to be steriliser is
considered sterile if, during the cycle, it passes through the sterilisation phase (PR). If this is not the case, it is consid-
ered "Load not sterile". In this case, restart the sterilisation cycle.
The sterilisation cycle can be stopped at any time by pressing the Stop button.
Cycle running
To stop the
cycle
4.6 Delayed start of cycle – overnight cycle
After selecting the cycle from the appropriate menu, the machine indicates the possibility of programming a delayed
start of the sterilisation cycle:
Cycle menu
To confirm
Up button
Down button
After selecting the cycle start option, enter your selection (if you wish the machine to start immediately, select
“
Immediate
”
, using the up and down arrows) by pressing the corresponding button.
If, on the other hand, you wish the machine to run during the night so that the equipment is ready to use in the morn-
ing, proceed as follows:
Enable the ventilation phase VEN (se section 4.3.7, page 15)
Set the delay to, for example, 10 hours: the surgery closes at 20:00 and reopens at 08:00, calculate 1 hour for
the sterilisation cycle, this gives you: from 20:00 to 08:00 is 12 hours, including the ventilation phase, the ma-
chine will start at 06:00 and will finish the sterilisation cycle at 07:00; the ventilation phase will then start to
prevent condensation inside the sterilisation chamber and cool the sterilised equipment.