FAST B
No
Message displayed
Description
Solution
13
CYCLE STOPPED BY OPE-
RATOR
Manual cycle stop
The cycle has been stopped by the operator: repeat the
loading and cycle start operations. If the problem per-
sists, contact technical assistance.
14
CHAMBER PRESSURE TOO
HIGH
Start pressure above atmospheric
Contact technical assistance.
15
TRIAC Q9 CLOSED
SUPPLY SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sists, contact technical assistance.
16
TEMPERATURE PHASE
PR HIGH
High temperature in phase PR
Allow the machine to cool. Inspect and clean the door
seal (maintenance section 5). Repeat the cycle. If the
problem persists, contact technical assistance.
17
TEMPERATURE PHASE
PR LOW
Low temperature in phase PR
Allow the machine to cool. Inspect and clean the door
seal (maintenance section 5). Repeat the cycle. If the
problem persists, contact technical assistance.
18
TRIAC Q10 CLOSED
SUPPLY SHUT OFF
Electronic component
Switch the machine off and on again. If the problem per-
sists, contact technical assistance.
19
TRIAC Q15 CLOSED
SUPPLY SHUT OFF
Electronic component
Switch the machine off and on again. If the problem per-
sists, contact technical assistance.
20
TRIAC Q13 CLOSED
SUPPLY SHUT OFF
Electronic component
Switch the machine off and on again. If the problem per-
sists, contact technical assistance.
21
FANS BLOCKED
Fans blocked
Contact technical assistance.
22
BIM. CHAMBER
Chamber heating element bimetallic safety
cut-out
Contact technical assistance.
23
BIM. STEAM GEN.
Steam generator heating element bimetal-
lic safety cut-out
Contact technical assistance.
24
SENSOR PTV CHAMBER
Chamber PTV sensor
Contact technical assistance.
25
SENSOR PTC CHAMBER
Chamber PTC sensor
Contact technical assistance.
26
SENSOR PTV
STEAM GENERATOR
Steam generator PTV sensor
Contact technical assistance.
27
DOOR MICROCONTACT
Door micro-switch
Contact technical assistance.
28
SENSOR CHAMBER PRES-
SURE
Pressure sensor
Contact technical assistance.
29
LOSS OF DATA E2P
Data loss E2P
Contact technical assistance.
30
DOOR MOTOR OR OPEN
MICROCONTACT
Door motor or open micro-switch
Contact technical assistance.
31
AL_TRIAC_Q8_CLS
Electronic component
Switch the machine off and on again. If the problem per-
sists, contact technical assistance.
32
NO MAINS DURING CYCLE
Power cut during cycle
Repeat the sterilisation cycle