4.8 Alarm messages and signals
Throughout the sterilisation cycle, the machine displays messages on the touch-screen. These inform the
operator of
the correct operation of the
appliance or any fault or error occurring during the sterilisation cycle (alarm).
The table below lists the alarms and respective messages programmed in the machine:
No. Message displayed
Description
Solution
1
FILL TANK
Tank fill level low
Top up demineralised water, section 3
2
FILL TANK TO BE EMP-
TIED
Tank full
Empty used water tank, section 3.2
3
MISSING LINE LOCK
No Sync signal (50 Hz)
Repeat the cycle. If the problem persists, contact tech-
nical assistance.
4
FILLER TANK SENSOR
DEFECTIVE
Filler tank sensor
Contact technical assistance
5
TRIAC Q8 OPEN
SUPPLY SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sists, contact technical assistance.
6
TRIAC Q9 OPEN
SUPPLY SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sists, contact technical assistance.
7
TRIAC Q10 OPEN
SUPPLY SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sists, contact technical assistance.
8
PRESSURE WRONG IN
THE PHASE PR
Alarm low temperature in phase PR
Allow the machine to cool. Inspect and clean the door
seal (maintenance section 5). Repeat the cycle. If the
problem persists, contact technical assistance.
9
WRONG TEMPERATURE IN
THE PHASE PR
Alarm, high temperature in phase PR
Allow the machine to cool. Inspect and clean the door
seal (maintenance section 5). Repeat the cycle. If the
problem persists, contact technical assistance.
10
PHASE NOT COMPLETED
Vacuum phase not executed
Dirt on door seal:
Inspect and clean the door seal (maintenance section 5).
Repeat the cycle. If the problem persists, contact tech-
nical assistance.
11
TANK TEMP. HIGH COO-
LING PHASE
Chamber heating element temperature
high: cooling pause
The sterilisation chambre temperature is high: wait for
cooling with door open
12
TRIAC Q15 OPEN
SHUT OFF
Electronic component
Turn the machine off and on again. If the problem per-
sist, contact technical assistance.