3-5 R eplacing s ens or probe
3 - 13
Replacing sensor probe
You can replace the sensor probes in the manner described below.
WARNING
Particular care should be paid to the keen-edged head of the probe. Otherwise it may be an injury to your
body.
1.
Select [Tool]
à
[Self diagnosis]
à
[Adjust probe]
à
[Probe 2] / [Probe 3] on the system and press [TEST
START] SW. (Probe moves to the front side
9
2.
Open the safe cover.
3.
Unscrew the hex-head screw, and remove the sensor probe.
4.
Disconnect a connector of the sensor probe.
5.
Install a new sensor probe as instructed below.
X-direction
A border line between green side and
white side should be in parallel with
the X-direction.
Direction of sensor probe
Sensor probe
Some space
There is some space
Hex-head screw
Connector
Sensor probe
CS-9000-A24
X-direction