![TAKAYA APT-9411 Series Service Manual Download Page 64](http://html2.mh-extra.com/html/takaya/apt-9411-series/apt-9411-series_service-manual_3316112064.webp)
R eplacing probe wire
3 - 11
Probe 1/4
You can replace the probe wires for Probe1/4 in the manner described below.
WARNING
Particular care should be paid to the keen-edged head of the probe to be replaced. Otherwise it may be an
injury to your body.
1.
Select [Tool]
à
[Self diagnosis]
à
[Adjust probe]
à
[Probe 1] / [Probe 4] on the system and press
[TEST START] SW. (Probe moves to the front side
9
2.
Open the Safe cover.
3.
Take the probe wire out of the PCB.
4.
Remove the probe holder and pull out the probe wire straight.
5.
Joint a new probe wire to the probe and install the probe holder.
(NOTE)
Probe wire for Probe 1/4 is TAKAYA LS-458-01 L=280mm.
6.
Install the probe wire up to the PCB in the direction which a figure below directs.
7.
Put two
"
Markers
attached on the probe wire between cable holders. (Two positions below)
8.
Move the probe holder up and down by hand to check neither the probe wire interferes nor stuck during
the movement.
9.
Close the Safe cover, then select [Close] and press [RESET] SW
à
[TEST STOP] SW.
10. Set the Machine Ref. Point again. (Refer to the APT-94xx Operator
#
s guide)
11. When the probe wire for Probe 4 was replaced, be sure to readjust the Camera offset. (Refer to the
APT-94xx Operator
#
s guide)
Probe wire
Screw
Cable holder (Two positions)
Probe holder
Probe
PCB
Cable holder
Marker
Probe wire
(Pay attention to the direction!)
(NOTE)
Never connect here!
Magnification of cable holder
3
45
#