Useful Function
------------------------------------------------------------------------------------
50
Measurement Display Smoothing
Use when wishing to stabilize displayed values when the load current is periodically fluctuating at high speeds.
The displayed value will be the average value of data, including new data, measured within the past 2 seconds.
This is renewed approximately every 200ms. (This method is known as the Moving Average Method)
Setting Procedures
Press the Function “FUNC.” key.
FUNC. lamp will light.
Settings Item Number will display on the top “Number
Display” and parameters on the bottom Number Display.
Select Item Number 12 with the VOLTAGE dial.
Set the parameter to 1 with the CURRENT dial.
Press the FUNC. to finish.
Returns Number Display to measurement display mode.