Using in Series/Parallel Operation
------------------------------------------------------------------------------------
127
2.Calibrating Slave Unit Output Settings
Calibrate the Slave unit to match the Slave’s output voltage with the Master’s voltage when in Series Operation.
Preparation
The following measuring device is required for calibration.
Voltage Measuring Digital Multimeter - 2 devices
No. of Display Digits: 5 1/2 or more Display Accuracy : 0.005% or more
※ (Equipped with a measuring range from 1mV~100V)
Offset/Full Scale Calibration of Slave Unit Output Voltage
1.
Detach the Load Wires and the Remote Sensing
Wires. Connect the Digital Multimeter to the output
terminal.
3. Press the PRESET key again and then the OUTPUT
key and set the output to ON.
Have the POWER switch turned ON and the output
turned OFF.
2.Press the PRESET key ① of the Master unit and
set to “0.01V” with the VOLTAGE dial ② and
“1.00A” with the CURRENT dial ③.
(See “Using as Constant Voltage Power Supply”
under “Basic Operations” on how to do this.)
4. Press the FUNC. key and turn the VOLTAGE dial to
display Settings Item 37 in the top Number Display.
5. Adjust the Slave unit's settings with the Current dial
to match the measured Master Unit's output voltage
displayed on the digital multimeter.
FUNC. lamp will light and the display in bottom Number Display
will disappear.
MEMO
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
①
②
+
-
V
+
-
V
スレーブ機
アナログコントロール
コネクタ
直列コントロール
コネクタ
マスター機
デジタルマル
チメータ
+
-
+
-
デジタルマル
チメータ
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
⇒Continued on next page
Analog Control
Connector
Series Control
Connector
Slave
Master
Digital
Multimeter
Digital
Multimeter
The ON/OFF of the "OUTPUT" is only valid with the Master
unit.
MEMO
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
DISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
③
ZX-S-400LN
Summary of Contents for ZX-S-LN Series
Page 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Page 13: ...About this Product 10 This is blank page...
Page 63: ...Useful Function 60 This is blank page...
Page 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Page 157: ...Special Loads 154 This is blank page...
Page 199: ...Specifications 196 This is blank page...
Page 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Page 201: ...Appendix 198 ZX S 800LN 800LAN...
Page 202: ...Appendix 199 ZX S 1600LN 1600LAN...