25
Customer Service Toll-Free: 866.435.3948
Technical Information
CHAPTER 9
Dimensions and Weights
Component
Size cm (inches)
Weight kg (lbs)
Control Unit
18.7 x 6.9 x 3.2 (7.4 x 2.7 x 1.3)
0.22 kg (0.49 lbs)
Operating Conditions
Temperature
+10°C to +40°C (+50°F to +104°F)
Humidity
0 to 75% RH
Pressure
0.7 to 1.3 Bar (70 kPa to 130 kPa)
Storage Conditions
Store at room temperature. Avoid excessive heat and cold.
Pressure Parameters
Mode
Foot
(± 5 mmHg)
Lower Calf
(± 5 mmHg)
Middle Calf
(± 5 mmHg)
Upper Calf
(± 5 mmHg)
Sustained Compression
40
40
30
20
Intermittent Pneumatic Compression
50
50
45
40
Equipment Classifications and Standards
The ACTitouch System is tested to/complies with the following equipment classifications and
standards:
U.S. Medical Equipment Classification
Class II
Degree of Protection Against Electric Shock
Class II
Classification According to Directive 93/42/EEC IIA
Safety
UL60601-1 and CAN/CSA C22.2 No.601.1–M90
Electromagnetic Compatibility (EMC)
EN60601-1-2
Software
EN60601-1-4
Internal Power Source
Lithium Ion Battery
External PSU Input
100–240Vac, 200mA, 50–60Hz, Class II
External PSU Output
7.5Vdc, 900mA
Summary of Contents for Actitouch