19
Customer Service Toll-Free: 866.435.3948
DisCide® ULTRA Spray Disinfectant has been demonstrated to effectively
disinfect the ACTitouch System. Use DisCide ULTRA Spray or similar disinfectant
compliant with OSHA’s Bloodborne Pathogen Standard (29 CFR 1910.1030) and/
or registered with EPA. To disinfect the ACTitouch control unit between patient
use, or if there are visible biological contaminants or visible stains, the following
steps are recommended:
1. Clean any visible blood or body fluids from the surface of the controller.
2. Thoroughly wet surface with DisCide ULTRA Disinfecting Spray.
3. Allow surfaces to remain wet for one minute and then allow to air dry.
Do not clean or disinfect with household detergents, abrasive cleaners,
scourers, degreasers, solvents, bleach or phenol-based agents, such as
trichlorophenol (TCP).
7 .2
Storage and Handling
It is recommended that the ACTitouch System be stored at room temperature.
Avoid excessive heat and cold. Do not store device in direct sunlight. Ensure that
the device is clean and dry prior to storage.
Reasonable care should be taken when handling and using the ACTitouch
System. Although the device has been designed for everyday use, heavy
impacts, contact with sharp objects and rough handling should be avoided. Do
not immerse the device in fluid. If either the Compression Sleeve or control unit
become soaked with fluid or otherwise damaged, do not use the device.
7 .3
Maintenance and Service
The ACTitouch System is designed to be maintenance free; routine service is not
required.
CAUTION: Do not open or take apart the ACTitouch control unit for any reason .
There are no customer serviceable parts in the ACTitouch System .
7 .4
Disposal
For disposal of any components of the ACTitouch System, please follow local
waste regulations or consult your local institutional waste-management service
or municipal waste authority.