Synthes
55
1
Insertion of the end cap
Instruments
03.010.110
Screwdriver Stardrive, T40, cannulated,
length 300 mm
357.399 Guide
Wire
3.2 mm
The end caps for the Expert Tibial Nails are available in exten-
sion lengths of 0 mm (04.004.000 and 04.004.004), 5 mm
(04.004.001), 10 mm (04.004.002), and 15 mm
(04.004.003). They fulfill three functions: they prevent bone
ingrowth into the nail; they extend the nail height if it is
overinserted; and they lock the proximal oblique screw or the
distal oblique locking screw, providing a stable, fixed-angle
construct.
The end caps are cannulated for use over a guide wire if nec-
essary.
Remove the nail insertion instruments.
To aid in end cap insertion, remove the connecting screw
only. The insertion handle can remain to help align the end
cap to the top of the nail. The end cap fits through the barrel
of the insertion handle.
Note:
The patient’s leg should be positioned in flexion to
facilitate end cap insertion.
Insert the end cap