40
Synthes
Expert Tibial Nail
5
Insert locking screw
Instrument
03.010.107
Screwdriver Stardrive, T25,
length 330 mm
Insert the appropriate locking screw through the protection
sleeve using the Stardrive T25 screwdriver. Verify locking
screw length under image intensification.
The tip of the locking screw should not project more than
1–2 mm beyond the far cortex.
Repeat the steps 3 to 5 for the second proximal ML locking
Screw.
Option
Additional cancellous bone locking screws can be added for
proximal fractures and highly unstable fractures.
Refer to steps 1 to 7 on pages 44 to 54 for details on proxi-
mal locking with the cancellous bone locking screws.
Proximal locking
0X6.000.380_AC.qxp:0X6.000.380_AC 20.01.2009 10:31 Uhr Seite 40