58
Synthes
Expert Tibial Nail
1
Remove end cap and locking screws
Instruments
03.010.110
Screwdriver Stardrive, T40, cannulated,
length 300 mm
03.010.107
Screwdriver Stardrive, T25,
length 330 mm
03.010.112
Holding Sleeve, with Locking Device
Implant removal is an optional procedure.
Clear the Stardrive socket of the end cap and the locking
from any tissue ingrowth. Remove the end cap with the
screwdriver Stardrive T40.
Remove all locking screws except one of the proximal locking
screws using the screwdriver Stardrive T25 and the holding
sleeve.
Note:
Always remove the most proximal cancellous bone
locking screw in order to insert the extraction screw into the
proximal end of the nail.
Implant removal
0X6.000.380_AC.qxp:0X6.000.380_AC 20.01.2009 10:31 Uhr Seite 58