DYNAM
STAT 2
STAT 1
Proximal locking
Nail Diameter
Locking Screws
Protection Sleeve
Drill Sleeve
Trocar
Calibrated Drill Bit
8 mm and 9 mm
4.0 mm
12.0 mm / 8.0 mm
8.0 mm / 3.2 mm
3.2 mm
3.2 mm
(dark blue)
(dark blue)
03.010.063
03.010.064
03.010.069
03.010.060
10 mm – 13 mm
5.0 mm
12.0 mm / 8.0 mm
8.0 mm / 4.2 mm
4.2 mm
4.2 mm
(light green)
(light green)
03.010.063
03.010.065
03.010.070
03.010.061
Diaphyseal and distal segment fractures
1
Choose locking screws and instruments
Use the correct locking screw, drill sleeve, trocar and drill bit
for the selected nail diameter as shown in the table.
Three proximal ML locking options can be targeted using
the aiming arm:
1
The dynamic locking option (DYNAM) corresponds to the
upper position of the proximal locking slot. This type
of locking allows primary compression or secondary, con-
trolled dynaminization of the bone fragments.
2
Static 2 (STAT 2) corresponds to the lower position of the
proximal locking slot. This type of locking does
not allow primary compression or secondary controlled dy-
naminization.
3
Static 1 (STAT 1) corresponds to most distal of the proximal
locking holes.