1
3
2
Air Power Line II Instructions for Use
DePuy Synthes
13
Technical data is subject to tolerances.
Quick Coupling for Guide Pins
B
1.5 to 4.0 mm
(510.790)
(Speed: 0–930 rpm)
Load the guide pin and insert into the bone
1. Adjust the adjusting sleeve
3
to fit the diameter of the
guide pin to be used. Slightly press the adjusting sleeve
axially against the tool and rotate the sleeve.
2. Apply a slight amount of pressure to insert the guide pin
from the front or back into the cannulation
1
. The guide
pin is held automatically.
3. Adjust the working length by pulling on the guide pin.
4. To fit the guide pin pull the tension lever
2
against the
tool. Only pull the tension lever against the tool as much
as necessary. The clamping force can be varied by pulling
and releasing the clamping lever.
5. Insert the guide pin into the bone. Apply the clamping
force as long as the guide pin is advanced.
6. To regrip the guide pin, reduce the clamping force and
move the tool to the desired length. Reclamp the guide
pin by pulling on the tension lever.
The guide pin is removed from the bone analogue as
described above.