34
DePuy Synthes
Air Power Line II Instructions for Use
Care and Maintenance
Precautions:
To ensure long service life and reduce repairs,
it is necessary that the accessible moving parts of the
handpiece and attachment are lubricated after each use.
Failing to lubricate the parts will lead to damage and mal-
function, increasing the risk of harm to the user and patient.
For further information on lubrication, please refer to the
Instruction for Use of the Synthes Special Oil 519.970
(60099544) and the APL II Care and Maintenance Poster
(038.000.017).
– Power tools and accessories may only be lubricated with
Synthes special oil (519.970). The composition of the
vapor-permeable and biocompatible oil is optimized for
the specific requirements of the power tools. Lubricants
with other compositions can cause the power tools to jam
and be toxic.
– Only lubricate the power tools and attachments when
clean.
– Oil should only be used on the air hose sleeve. Ensure that
no oil gets into the double air hose. When lubricating,
never use a double hose without the lubrication adapter
(519.790) since leaking oil could otherwise damage the
hose.
Warning:
The Radiolucent Drive (511.300) does not require
lubrication.