Air Power Line II Instructions for Use
DePuy Synthes
37
Precautions:
– The following maximum values may not be exceeded:
138°C over a maximum of 18 minutes. Higher values can
damage the sterilized products.
– Observe the packages prior to storage for visual moisture
or dampness and if found on or within the pack, the
product should be repackaged and sterilized with an in-
creased drying time.
– Do not accelerate the cooling process as it will damage
the electronic components of the power tool and could
result in harm to the user and patient.
– Hot air, ethylene oxide, plasma and formaldehyde steriliza-
tion are not recommended.
– Ensure that the seal nipple (519.596) is removed from the
machine’s air inlet and that the air hose’s inlet and outlet
are disconnected prior to sterilization.
Storage
Storage conditions for products labeled “STERILE” are
printed on the packaging label.
Packaged and sterilized products should be stored in a dry,
clean environment, protected from direct sunlight, pests,
and extremes of temperature and humidity. Use products in
the order in which they are received (“first-in, first-out princi-
ple“), taking note of any expiration date on the label.