Air Power Line II Instructions for Use
DePuy Synthes
29
Automated Cleaning Instructions
with Manual Pre-cleaning
Manual pre-cleaning prior to automated cleaning/disinfec-
tion is important to ensure cannulations and other difficult
to access areas are clean.
Alternative cleaning/disinfection procedures other than in the
procedure described below (including manual pre-cleaning)
have not been validated by Synthes.
Preparation for manual cleaning and automated cleaning
with manual pre-cleaning:
Article number
Procedure
Air Power Line II handpieces
Put the seal nipple (519.596) on the
machine’s air inlet.
511.606
511.610
511.615
Synthes Double Air Hoses
Seal the Synthes Double Air Hose by
joining the inlet and outlet.
519.510
519.530
519.550
Dräger Double Air Hoses
Connect both sides of the Dräger Double
Air Hose with the seal nipple (519.596).
519.610
519.630
519.650
BOC/Schrader Double Air Hoses
Connect both sides of the BOC/Schrader
Air Hose with the seal nipple (519.591 or
519.592).
519.511
519.531
Note:
Before connecting the air inlet of the handpiece and
both sides of the air hose with the seal nipple, make sure
that the surfaces, which the seal nipple will cover, are not
contaminated. If they are, first wipe off or spray these
surfaces with alcohol-based disinfectant and then put the
seal nipple on. Make sure that no solution enters the hose.
Precaution:
Clean all movable parts in opened or unlocked
position.