DePuy Synthes
Air Power Line II Instructions for Use
Care and Maintenance
4. Clean with detergent
Clean the device manually under running warm water using
an enzymatic cleaner or detergent for a minimum of 5 min-
utes. Manipulate all moving parts under running water. Use
a soft-bristled brush and/or soft lint-free cloth to remove all
visible soil and debris. Follow the enzymatic cleaner or deter-
gent manufacturer’s instructions for use for correct tempera-
ture, water quality and concentration/dilution.
5. Rinse with tap water
Rinse the device thoroughly using cool to lukewarm running
water for a minimum of 2 minutes. Use a syringe or pipette
to flush lumens and channels. Actuate joints, handles and
other movable device features in order to rinse thoroughly
under running water.
6. Visually inspect device
Inspect the cannulations, sliding sleeves, attachment release
rings, etc. for visible soil. Repeat steps 1–6 if visible soil re-
mains.
7. Final rinse with deionized/purified water
Final rinse with deionized or purified water for a minimum of
2 minutes.
8. Dry
Dry device using a clean, soft lint-free cloth or medical grade
compressed air.