Air Power Line II Instructions for Use
DePuy Synthes
3
General Information
For the tool to function properly, Synthes recommends that
it is cleaned and serviced after each use in accordance with
the process defined in the “Care and Maintenance” section.
Compliance with these specifications can considerably ex-
tend the service life of the tool and reduce the risk of mal-
function or harm to the user and patient. Only use Synthes
Special Oil (519.970) to lubricate the tool.
We recommend using new Synthes cutting tools for every
surgical procedure. Efficiently working cutting tools are the
basis for successful surgery. Therefore, check used cutting
tools after every use for wear and/or damage and replace
them if necessary.
Cutting tools must be cooled with irrigation fluid to prevent
heat necrosis.
Unusual Transmissible Pathogens
Surgical patients identified as at risk for Creutzfeldt-Jakob
disease (CJD) and related infections should be treated with
single-use instruments. Dispose of the instruments used or
suspected of use on a patient with CJD after surgery and/or
follow the current national recommendations.
Servicing
This system requires regular maintenance service, at least
once a year, in order to maintain its functionality. This service
has to be performed by the original manufacturer or an au-
thorized site.
The manufacturer assumes no responsibility for damage re-
sulting from improper operation, neglected or unauthorized
maintenance of the tool.
Intended use
The Air Power Line II is a compressed air powered system to
be used for treatment in orthopedic and traumatology sur-
gery, i.e. drilling, reaming, cutting, placing of Kirschner Wires
on bone of human skeleton.
Safety Instructions
The surgeon has to evaluate if the machine is suitable for an
application, based on power limitation of the machine, at-
tachment and cutting tool regarding bone strength/anatomi-
cal situation as well as handling of the machine, attachment
and cutting tool regarding bone size. In addition, the contra-
indications of the implant have to be respected. Please refer
to the corresponding “Surgical Techniques” of the implant
system used.
The Air Power Line System is only to be used for patient
treatment after careful consultation of the instructions for
use. It is recommended that an alternative system is available
to use during application, as technical problems can never be
completely ruled out.
The Air Power Line System is designed for use by physicians
and trained medical personnel.
DO NOT use any apparently damaged components.
DO NOT use any component if the packaging is damaged.
To ensure correct operation of the Air Power Line, use only
Synthes original accessories.
Recommended operating pressure: 6–7 bar (max 10 bar).
Only use original Synthes hoses for compressed air.
Before the first and every use and prior returning for service,
power tools and their accessories/attachments have to run
through the complete reprocessing procedure. Protective
covers and films must be fully removed before sterilization.
The user of the product is responsible for proper user of the
equipment during surgery.
Check correct operation of the tools before using it on the
patient.
High torque of the powerful Reamer/Drill (511.606) must
always be observed.
Introduction