Operating Instructions
27
In the "
Setup
" main menu point the various sub-menu points should be selected in sequence and the
appropriate parameters entered for them to ensure that the measured is set perfectly.
The procedure is described below.
Setup / Medium
Each medium has different reflection properties. Other factors for solid include the dust development, material
cone and additional echoes from the vessel wall.
The "
Solid
" selection should be made first in this menu point to adjust the sensor to these different
measurement conditions.
This selection will adjust the sensor perfectly to the product and significantly increase measurement accuracy,
primarily for media with poor reflection properties.
Enter the required parameters using the appropriate keys, save them by pressing
[OK]
and then go to the next
menu point by pressing
[ESC]
and
[->]
.
Setup / Application
In addition to the medium the application or place of use can also affect the measurement.
This menu point enables you to adjust the sensor to the measuring conditions.
The following selections are available for "
Solid
":
The applications are based on the following features:
Silo (slim and tall):
•
Metal vessel: Welds
•
Process / Measuring conditions
– Filling close to the sensor
– System noise increased when the silo is completely empty
– Automatic interference signal hiding for partly filled vessel