10.4.50
SetDetorCom
Sets the bias voltage on the detor. The bias voltage is related to the 8 bit code by the formula;
Volts = (code/255) * 4.096V * 0.75
Command
Packet Byte
Contents
0
0x02 - STX
1
0x01 - LSB (message size – 1)
2
0x00 - MSB (message size – 1)
3
0x16 - Command Byte LSB
4
0x00 - Command Byte MSB
5
bias voltage code
6
0x00
7
0x00 - Unused
Echo
Packet Byte
Contents
0
0x02 - STX
1
0x01 - LSB (message size – 1)
2
0x00 - MSB (message size – 1)
3
0x00 - (unused message byte)
4
0x00 - (unused message byte)
5
0x00 - Unused
6
0x00 - Unused
10.4.51
GetDetorCom
Returns the value of the detor bias voltage.
Command
Packet Byte
Contents
0
0x02 - STX
1
0x01 - LSB (message size – 1)
2
0x00 - MSB (message size – 1)
3
0x16 - Command Byte LSB
4
0x01 - Command Byte MSB
5
0x00 (unused message byte)
6
0x00 (unused message byte)
7
0x00 - Unused
Echo
Packet Byte
Contents
0
0x02 - STX
1
0x01 - LSB (message size – 1)
2
0x00 - MSB (message size – 1)
3
bias voltage code
4
0x00 - Unused
5
0x00 - Unused
Summary of Contents for Acuros CQD Series
Page 13: ...User Manual Acuros CQD Camera 13 P a g e Figure 3 Acuros CQD 1920 GigE with F mount...
Page 14: ...User Manual Acuros CQD Camera 14 P a g e Figure 4 Acuros CQD 1280 USB with C mount...
Page 15: ...User Manual Acuros CQD Camera 15 P a g e Figure 5 Acuros CQD 1280 GigE with C Mount...
Page 16: ...User Manual Acuros CQD Camera 16 P a g e Figure 6 Acuros CQD 640 USB with C Mount...
Page 92: ...User Manual Acuros CQD Camera 92 P a g e...
Page 101: ...User Manual Acuros CQD Camera 101 P a g e 5 0x00 Unused 6 0x00 Unused...
Page 108: ...User Manual Acuros CQD Camera 108 P a g e 8 0x00 Unused...
Page 110: ...User Manual Acuros CQD Camera 110 P a g e 6 0x00 Unused...