AMU pH-Redox
Installation
A-96.250.441 / 060619
31
3.7.
Install Spray Nozzle (Option) into M-Flow Cell
To install the optional spray nozzle proceed as follows:
1
Stop operation according to chapter
2
Remove the sensors according to chapter 6, paragraph
move electrodes from flow cell, p. 45
.
3
Remove the calibration vessel [H] from the flow cell block [F]
and empty it.
4
Unscrew and remove the sealing screw form the cleaning solu-
tion inlet [B].
5
Insert the spray nozzle [G] so that its pin fits into the guiding slot
of the cleaning solution inlet.
6
To fix the spray nozzle screw the enclosed M4 screw into the
threaded hole [C] next to the cleaning solution inlet.
7
Fix the calibration vessel to the flow cell block.
8
Unscrew and remove the blind plug [D].
9
Install the hose nozzle [E].
10
Install the sensors.
A
B
C
D
Flow cell block bottom view
Cleaning solution inlet
Threaded hole for fixing screw
Blind plug
E
F
G
H
Hose nozzle
Flow cell block
Spray nozzle
Calibration vessel
open
close
B C
D
F
G
H
E
A