AMU pH-Redox
Program List and Explanations
A-96.250.441 / 060619
65
5 Installation
5.1 Sensors
5.1.1
Flow
5.1.1.1
Flow measurement
: Select “Q-Flow” or “None”. The deltaT flow
sensor is not supported by the AMU transmitter.
5.1.2
Parameters:
5.1.2.1
Type of Sensor
: Set the type of installed sensor. Possible types:
5.1.2.2
Sensor Check
: Default setting is: off. (
Only visible in pH measuring
mode
).
The <Sensor Check> can be enabled for automatic, periodic check
of the sensor aging (measuring the impedance). The measuring
values are written in the logger file or transferred to a control room
via Profibus or Modbus. It does not indicate an Alarm.
5.1.3
Temperature:
5.1.3.1
Temp. Sensor
: The pH measurement is temperature dependent
therefore it is possible to install a temperature sensor. Depending
on your configuration set the Temp. Sensor to:
If No is set the measuring value is compensated with the default
temperature.
5.1.3.21
Default Temp
.: If no temperature sensor is installed, set the default
temperature to the assumed average temperature of the sample.
The measuring value is then compensated with this value.
Flow measurement
Q-Flow
None
Q-Flow
deltaT
Type of Sensor
pH
Measuring values are displayed as pH
Redox
Measuring values are displayed in mV
Sensor check
off
on
Temp. Sensor
Yes
No